| Naslov: | Intra‐rater and inter‐rater reliability of robotic arthrometer DYNEELAX® |
|---|
| Avtorji: | ID Mihalinec, Katja (Avtor) ID Martinez-Cepa, Carmen B. (Avtor) ID Zuil-Escobar, Juan C. (Avtor) ID Kejžar, Nataša (Avtor) ID Vauhnik, Renata (Avtor) |
| Datoteke: | PDF - Predstavitvena datoteka, prenos (171,19 KB) MD5: DC72B485C9403184ED142349C0394747
URL - Izvorni URL, za dostop obiščite https://esskajournals.onlinelibrary.wiley.com/doi/10.1002/jeo2.70026
|
|---|
| Jezik: | Angleški jezik |
|---|
| Tipologija: | 1.01 - Izvirni znanstveni članek |
|---|
| Organizacija: | UKC LJ - Univerzitetni klinični center Ljubljana
|
|---|
| Povzetek: | Purpose: The purpose of our study was to test the intra-rater and inter-rater reliability of the DYNEELAX® arthrometer in healthy subjects. Since rotational laxity will be measured for the first time in humans, indications for normative values will also be presented. Methods: Knee anterior laxity and rotational laxity using a DYNEELAX® arthrometer were assessed in 73 subjects (39 females and 34 males). An intraclass correlation coefficient was calculated to evaluate the intra-rater and inter-rater reliability of the DYNEELAX®. Results: An intraclass correlation coefficient for intra-rater reliability ranges from 0.631 (95% confidence interval; [CI]: 0.47–0.75) to 0.699 (95% CI: 0.56–0.80) and from 0.916 (95% CI: 0.87–0.95) to 0.94 (95% CI: 0.91–0.96) for anterior knee laxity and rotational knee laxity, respectively. An intraclass correlation coefficient for inter-rater reliability ranges from 0.578 (95% CI: 0.40–0.71) to 0.646 (95% CI: 0.44–0.78) and from 0.822 (95% CI: 0.57–0.91) to 0.933 (95% CI: 0.89–0.96) for anterior knee laxity and rotational knee laxity, respectively. Conclusions: The DYNEELAX® arthrometer has good to excellent intra-rater and inter-rater reliability for rotational knee laxity and moderate intra-rater reliability for anterior knee laxity in healthy subjects. Future studies should investigate the clinical significance of anterior and rotational laxity measured with the DYNEELAX® arthrometer in patients with knee pathology, as both laxities are critical for assessing the integrity of the intra-articular structures of the knee in clinical practice. |
|---|
| Ključne besede: | anterior laxity, arthrometry, rotational laxity, reliability |
|---|
| Status publikacije: | Objavljeno |
|---|
| Verzija publikacije: | Objavljena publikacija |
|---|
| Leto izida: | 2024 |
|---|
| Št. strani: | str. 1-9 |
|---|
| Številčenje: | Vol. 11, issue 4, [article no.] e70026 |
|---|
| PID: | 20.500.12556/DiRROS-29177  |
|---|
| UDK: | 615.82/.84:616.728.3 |
|---|
| ISSN pri članku: | 2197-1153 |
|---|
| DOI: | 10.1002/jeo2.70026  |
|---|
| COBISS.SI-ID: | 209810179  |
|---|
| Opomba: | Nasl. z nasl. zaslona;
Opis vira z dne 2. 10. 2024;
|
|---|
| Datum objave v DiRROS: | 23.04.2026 |
|---|
| Število ogledov: | 60 |
|---|
| Število prenosov: | 47 |
|---|
| Metapodatki: |  |
|---|
|
:
|
Kopiraj citat |
|---|
| | | | Objavi na: |  |
|---|
Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše