Digitalni repozitorij raziskovalnih organizacij Slovenije

Izpis gradiva
A+ | A- | Pomoč | SLO | ENG

Naslov:MRI-assisted cervix cancer brachytherapy pre-planning, based on application in paracervical anaesthesia : final report
Avtorji:ID Petrič, Primož (Avtor)
ID Hudej, Robert (Avtor)
ID Hanuna, Omar (Avtor)
ID Marolt, Primož (Avtor)
ID Šegedin, Barbara (Avtor)
Datoteke:URL URL - Izvorni URL, za dostop obiščite http://www.degruyter.com/view/j/raon.2014.48.issue-3/raon-2014-0009/raon-2014-0009.xml?format=INT
 
.pdf PDF - Predstavitvena datoteka, prenos (753,88 KB)
MD5: 1371E1940179B141C4E5EBE45DDCFDDD
 
Jezik:Angleški jezik
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:Logo OI - Onkološki inštitut Ljubljana
Povzetek:Background. Optimal applicator insertion is a precondition for the success of cervix cancer brachytherapy (BT). We aimed to assess feasibility and efficacy of MRI-assisted pre-planning, based on applicator insertion in para-cervical anaesthesia (PCA). Patients and methods. Five days prior to BT, the pre-planning procedure was performed in 18 cervix cancer patients: tandem-ring applicator was inserted under PCA, pelvic MRI obtained and applicator removed. Procedure tolerability was assessed. High risk clinical target volume (HR CTV) and organs at risk were delineated on the pre-planning MRI, virtual needles placed at optimal positions, and dose planning performed. At BT, insertion was carried out in subarachnoidal anaesthesia according to pre-planned geometry. Pre-planned and actual treatment parameters were compared. Results. Pre-planning procedure was well tolerated. Median difference between the pre-planned and actual needle insertion depth and position were 2 (0%10) mm and 4 (0%30) degrees, respectively. The differences between the pre-planned and actual geometric and dosimetric parameters were statistically non-significant. All actual needles were positioned inside the HR CTV and outside the organs at risk (OAR). Conclusions. Our pre-planning approach is well tolerated and effective. Pre-planned geometry and dose distribution can be reproduced at BT.
Ključne besede:cervix cancer, pre-planning, image-guided brachytherapy
Status publikacije:Objavljeno
Verzija publikacije:Objavljena publikacija
Datum objave:01.09.2014
Založnik:Association of Radiology and Oncology
Leto izida:2014
Št. strani:str. 293-300
Številčenje:Vol. 48, no. 3
Izvor:Ljubljana
PID:20.500.12556/DiRROS-18667 Novo okno
UDK:618.146-006:615.849
ISSN pri članku:1318-2099
DOI:10.2478/raon-2014-0009 Novo okno
COBISS.SI-ID:1853819 Novo okno
Avtorske pravice:by Authors
Datum objave v DiRROS:11.04.2024
Število ogledov:97
Število prenosov:40
Metapodatki:XML RDF-CHPDL DC-XML DC-RDF
:
Kopiraj citat
  
Objavi na:Bookmark and Share


Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše podrobnosti ali sproži prenos.

Gradivo je del revije

Naslov:Radiology and oncology
Skrajšan naslov:Radiol. oncol.
Založnik:Slovenian Medical Society - Section of Radiology, Croatian Medical Association - Croatian Society of Radiology
ISSN:1318-2099
COBISS.SI-ID:32649472 Novo okno

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:rak materničnega vratu, brahiterapija, slikovno vodena brahiterapija, magnetnoresonančno slikanje, MRI, vnaprejšnje načrtovanje


Nazaj