Digital repository of Slovenian research organisations

Show document
A+ | A- | Help | SLO | ENG

Title:Impact of CT tube voltage and anode current on PET/CT image quality : a nema IQ phantom study
Authors:ID Čavlin, Milana (Author)
ID Ležaič, Luka (Author)
ID Rep, Sebastijan (Author)
Files:.pdf PDF - Presentation file, download (762,36 KB)
MD5: 5BB82754A4ACBB1ABBCDFCA195BB5E88
 
URL URL - Source URL, visit https://doi.org/10.2298/NTRP2602152C
 
Language:English
Typology:1.01 - Original Scientific Article
Organization:Logo UKC LJ - Ljubljana University Medical Centre
Abstract:Positron emission tomography/computer tomography is a crucial modality for the quantification of metabolic activity and anatomical structure. Reducing radiation dose, especially in repeated imaging, remains a significant challenge. This study focuses on optimizing computer tomography parameters to reduce dose, while maintaining image quality. The aim of this research was to determine the impact of varying computer tomography tube voltage and anode current, on the quantification of standardized uptake value, contrast-to-noise ratio, and radiation exposure (volume CT dose index, dose-length product). A NEMA IQ body phantom was scanned using various combinations of tube voltage and anode current. Standardized up-take value, contrast-to-noise ratio, volume CT dose index and dose-length product, were evaluated from the different scans. The obtained results were statistically assessed. Reducing tube voltage and anode current did not significantly affect standardized uptake value accuracy but, reduced radiation dose. Image quality (contrast-to-noise ratio) remained diagnostically acceptable, supporting the potential for optimization without compromising accuracy. Optimization of computer tomography parameters (tube voltage, anode current) is feasible and beneficial in clinical practice, as it reduces patient radiation burden while maintaining image quality and standardized uptake value accuracy.
Keywords:PET/CT, standardized uptake value, quantification, dose optimization, image quality, tube voltage, anode current
Publication status:Published
Publication version:Version of Record
Year of publishing:2026
Number of pages:str. 152-158
Numbering:Vol. 41, iss. 2
PID:20.500.12556/DiRROS-31682 New window
UDC:61
ISSN on article:1452-8185
DOI:10.2298/NTRP2602152C New window
COBISS.SI-ID:286409987 New window
Copyright:Na spletni strani revije navedeno: Nuclear Technology & Radiation Protection is an open access Journal. All articles can be downloaded free of charge and used in accordance with the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 Serbia (https://ntrp.vinca.rs/Editorial%20policy.htm)
Note:Nasl. z nasl. zaslona; Opis vira z dne 30. 7. 2026; Povzetek v angl. in srb.;
Publication date in DiRROS:06.08.2026
Views:126
Downloads:85
Metadata:XML DC-XML DC-RDF
:
Copy citation
  
Share:Bookmark and Share


Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Record is a part of a journal

Title:Nuclear technology & radiation protection
Shortened title:Nucl. Techn. & Rad. Protect.
Publisher:Vinča Institute of Nuclear Sciences
ISSN:1452-8185
COBISS.SI-ID:526273305 New window

Licences

License:CC BY-NC-ND 3.0, Creative Commons Attribution-NonCommercial-NoDerivs 3.0 Unported
Link:http://creativecommons.org/licenses/by-nc-nd/3.0/
Description:You are free to reproduce and redistribute the material in any medium or format. You must give appropriate credit, provide a link to the license, and indicate if changes were made. You may do so in any reasonable manner, but not in any way that suggests the licensor endorses you or your use. You may not use the material for commercial purposes. If you remix, transform, or build upon the material, you may not distribute the modified material. You may not apply legal terms or technological measures that legally restrict others from doing anything the license permits.

Back