Digital repository of Slovenian research organisations

Show document
A+ | A- | Help | SLO | ENG

Title:International clinical opinion on transparency, standardisation, and calibration alignment in the performance evaluation of systems for continuous glucose monitoring
Authors:ID Pemberton, John S. (Author)
ID Andrews, Robert C. (Author)
ID Barnard-Kelly, Katharine (Author)
ID Battelino, Tadej (Author)
ID Danne, Thomas (Author)
ID Heinemann, Lutz (Author)
ID Kar, Partha (Author), et al.
Files:.pdf PDF - Presentation file, download (2,01 MB)
MD5: 3BFC416FA32E1BFFF8EFAAE2B11FF802
 
URL URL - Source URL, visit https://dom-pubs.pericles-prod.literatumonline.com/doi/10.1111/dom.70460
 
Language:English
Typology:1.02 - Review Article
Organization:Logo UKC LJ - Ljubljana University Medical Centre
Abstract:Continuous glucose monitoring (CGM) is now central to diabetes management, yet variation in how respective medical products are evaluated limits meaningful comparison between CGM systems. Three barriers currently constrain reliable interpretation of glucose-derived measures. The first is limited transparency: in several regulatory settings, particularly those using Conformité Européenne marking, clinical-study reports, reference-method information and analytical documentation required for market authorisation are not publicly accessible. The second barrier is heterogeneity in study procedures. Existing evaluations use different reference-glucose methods, sampling strategies, glucose-manipulation protocols and participant characteristics, leading to accuracy estimates that cannot be interpreted consistently across systems. The third barrier is calibration alignment. Even with full transparency and aligned procedures, CGM systems may differ because their calibration algorithms are trained on distinct reference-glucose datasets, influencing reported glucose ranges, automated insulin-delivery behaviour and interpretation during device transitions. A modified Delphi process involving clinicians, laboratory scientists, and researchers identified these issues as the principal determinants of comparability. During this process, the International Federation of Clinical Chemistry and Laboratory Medicine released a validated framework for performance evaluation of CGM systems, providing a unified approach to reference-method selection, dynamic in-clinic testing, and structured reporting. Adoption would reduce procedural variability but does not resolve calibration-alignment differences. This international clinical opinion proposes a pathway towards internationally interpretable CGM evaluation: immediate transparency of clinical evidence, routine declaration of calibration alignment, and progressive adoption of validated standardised procedures. These steps provide a foundation for reliable interpretation and globally comparable assessment of CGM technologies.
Keywords:continuous glucose monitoring, CGM, diabetes management, evaluation
Publication status:Published
Publication version:Version of Record
Year of publishing:2026
Number of pages:str. 2551-2565
Numbering:Vol. 28, issue 4
PID:20.500.12556/DiRROS-28435 New window
UDC:616.379
ISSN on article:1463-1326
DOI:10.1111/dom.70460 New window
COBISS.SI-ID:268352003 New window
Note:Nasl. z nasl. zaslona; Opis vira z dne 13. 2. 2026;
Publication date in DiRROS:19.03.2026
Views:205
Downloads:146
Metadata:XML DC-XML DC-RDF
:
Copy citation
  
Share:Bookmark and Share


Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Record is a part of a journal

Title:Diabetes, obesity & metabolism
Shortened title:Diabetes obes. metab.
Publisher:Blackwell Science
ISSN:1463-1326
COBISS.SI-ID:517693465 New window

Licences

License:CC BY-NC 4.0, Creative Commons Attribution-NonCommercial 4.0 International
Link:http://creativecommons.org/licenses/by-nc/4.0/
Description:A creative commons license that bans commercial use, but the users don’t have to license their derivative works on the same terms.