Digital repository of Slovenian research organisations

Show document
A+ | A- | Help | SLO | ENG

Title:Force signal alignment in dynamic testing machine calibration applications
Authors:ID Hiti, Miha (Author)
Files:URL URL - Source URL, visit https://doi.org/10.1016/j.measurement.2025.119070
 
.pdf PDF - Presentation file, download (6,07 MB)
MD5: 44368DB94C947995B5E891DAE37757BC
 
Language:English
Typology:1.01 - Original Scientific Article
Organization:Logo ZAG - Slovenian National Building and Civil Engineering Institute
Abstract:Calibration of force in measuring systems where the force is time dependent i.e. dynamically changing, introduces additional complexities versus quasi-static calibration conditions. Time variable force signals necessitate accurate time alignment between concurrent signals – measurement series from unit under test and the measurement series from the reference standard need to be synchronised. Different unsynchronised measuring systems are typically used for acquisition of each calibration signal and they need to be time aligned in post- processing to eliminate as much as possible the influence of the sample misalignment from the resulting testing machine indication error. The paper focuses on the time synchronization of measurement series by using cross-correlation to determine the necessary time correction when calibrating testing machines with dynamic cyclic force excitation. Measurements are presented for calibration signals in the examples of 20 kN testing machine compression calibration for 5 kN ± 1 kN cycling force with 1 Hz, 5 Hz, and 10 Hz cycling frequencies, and 1000 kN testing machine tensile calibration for 500 kN ± 100 kN cycling force with 1 Hz cycling frequency. The standard deviation of the error distribution after time correction and clock rate compensation was below 0.2 % for both, 20 kN testing machine calibration, and 1000 kN testing machine calibration.
Keywords:calibration, dynamic testing systems, time-synchronization, cross-correlation
Publication status:Published
Publication version:Version of Record
Publication date:20.09.2025
Publisher:Elsevier
Year of publishing:2026
Number of pages:str. 1-11
Numbering:Vol. 258, pt. A, [article no.] 119070
PID:20.500.12556/DiRROS-27523 New window
UDC:544.5/.6
ISSN on article:1873-412X
DOI:10.1016/j.measurement.2025.119070 New window
COBISS.SI-ID:267418371 New window
Copyright:© 2025 Slovenian National Building and Civil Engeneering Institute (ZAG)
Publication date in DiRROS:11.02.2026
Views:379
Downloads:107
Metadata:XML DC-XML DC-RDF
:
Copy citation
  
Share:Bookmark and Share


Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Record is a part of a journal

Title:Measurement
Publisher:Elsevier
ISSN:1873-412X
COBISS.SI-ID:23272709 New window

Document is financed by a project

Funder:Other - Other funder or multiple funders
Funding programme:European Metrology Programme for Innovation and Research
Project number:18SIB08
Name:Comprehensive traceability for force metrology services
Acronym:ComTraForce

Funder:ARIS - Slovenian Research and Innovation Agency
Project number:P2-0273-2019
Name:Gradbeni objekti in materiali

Licences

License:CC BY 4.0, Creative Commons Attribution 4.0 International
Link:http://creativecommons.org/licenses/by/4.0/
Description:This is the standard Creative Commons license that gives others maximum freedom to do what they want with the work as long as they credit the author.

Secondary language

Language:Slovenian
Keywords:kalibraicja, dinamični preskusni sistemi, časovna sinhronizacija, križan korelacija


Back