Naslov:Verification, implementation and harmonization of automated chemiluminescent immunoassays for MPO- and PR3-ANCA detection
Avtorji:ID Ogrič, Manca (Avtor)
ID Švec, Tinka (Avtor)
ID Mrak Poljšak, Katjuša (Avtor)
ID Žigon, Polona (Avtor)
ID Hočevar, Alojzija (Avtor)
ID Čučnik, Saša (Avtor)
Datoteke:.pdf PDF - Predstavitvena datoteka, prenos (514,85 KB)
MD5: E6E9F0F9C540CDE90EF8D1FF1CF177C2
 
URL URL - Izvorni URL, za dostop obiščite https://www.degruyter.com/document/doi/10.1515/cclm-2023-0764/html
 
Jezik:Angleški jezik
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:Logo UKC LJ - Univerzitetni klinični center Ljubljana
Povzetek:Objectives: Antineutrophil cytoplasmic antibody (ANCA) testing assists clinicians diagnose ANCA-associated vasculitis (AAV). We aimed to verify and harmonize chemiluminescent immunoassays for the detection of myeloperoxidase (MPO)- and proteinase 3 (PR3)-ANCA. Methods: An in-house ELISA, a capture ELISA, and a chemiluminescent assay QUANTA Flash on a BIO-FLASH analyzer were used to detect MPO- and PR3-ANCA in sera from 39 patients with AAV, 55 patients with various non-AAV, and 66 patients with connective tissue diseases. The results of the assays were evaluated, and their clinical performance was assessed. The precision and linearity of the QUANTA Flash assays were determined, and likelihood ratios (LRs) for AAV at diagnosis were calculated. Results: The precision and linearity of the QUANTA Flash assays were confirmed. Overall agreement between 97.5 and 98.8 % and Cohen's kappa coefficients between 0.861 and 0.947 were observed for the results of the QUANTA Flash assays and ELISAs. The diagnostic sensitivity, specificity, and ROC analysis of the assays for AAV were statistically similar (in-house ELISA 89.7 %, 95.0 %, and 0.937; capture ELISA 92.3 %, 98.3 %, and 0.939; and QUANTA Flash 89.7 %, 95.9 %, and 0.972). For the QUANTA Flash assay results, the interval-specific LRs for AAV at diagnosis were: 0-8 CU had LR 0.08, 8-29 CU had LR 1.03, 29-121 CU had LR 7.76, 121-191 CU had LR 12.4, and >191 CU had LR ∞. Conclusions: The QUANTA Flash MPO and PR3 assays provide precise and consistent results and have comparable clinical utility for AAV. The calculated LRs were consistent with published LRs, confirming the utility of LRs for harmonization of ANCA results.
Ključne besede:antineutrophil cytoplasmic antibodies (ANCA), proteinase 3 (PR3), myeloperoxidase (MPO), ANCA-associated vasculitis (AAV), chemiluminescent immunoassay, likelihood ratios (LRs)
Status publikacije:Objavljeno
Verzija publikacije:Objavljena publikacija
Leto izida:2023
Št. strani:str. 682-689
Številčenje:Vol. 62 [article no.] 0764, iss. 4
PID:20.500.12556/DiRROS-32092 Novo okno
UDK:616-097
ISSN pri članku:1437-4331
DOI:10.1515/cclm-2023-0764 Novo okno
COBISS.SI-ID:169691907 Novo okno
Opomba:Nasl. z nasl. zaslona; Opis vira z dne: 24. 10. 2023;
Datum objave v DiRROS:27.08.2026
Število ogledov:141
Število prenosov:82
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
  
Objavi na:Bookmark and Share



Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše podrobnosti ali sproži prenos.

Gradivo je del revije

Naslov:Clinical chemistry and laboratory medicine
Skrajšan naslov:Clin. chem. lab. med.
Založnik:de Gruyter
ISSN:1437-4331
COBISS.SI-ID:680355 Novo okno

Licence

Licenca:CC BY 4.0, Creative Commons Priznanje avtorstva 4.0 Mednarodna
Povezava:http://creativecommons.org/licenses/by/4.0/deed.sl
Opis:To je standardna licenca Creative Commons, ki daje uporabnikom največ možnosti za nadaljnjo uporabo dela, pri čemer morajo navesti avtorja.

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:antinevtrofilna citoplazemska protitelesa (ANCA), proteinaza 3 (PR3), mieloperoksidaza (MPO), vaskulitis, povezan z ANCA (AAV), kemiluminiscentni imunski test, razmerja verjetnosti (LR)


Nazaj