| Naslov: | Prevention of air embolism in extracorporeal membrane oxygenation systems : an in vitro study on protection of central venous catheter lumen |
|---|
| Avtorji: | ID Franco, Danilo (Avtor) ID Krašna, Nejc (Avtor) ID Novak, Robert (Avtor) ID Esposito, Giovanni (Avtor) ID Izzo, Raffaele (Avtor) ID Bělohlávek, Jan (Avtor) ID Noč, Marko (Avtor) ID Goslar, Tomaž (Avtor) |
| Datoteke: | PDF - Predstavitvena datoteka, prenos (1,59 MB) MD5: 07185AACDAAED668ECAD72C75DF207F6
URL - Izvorni URL, za dostop obiščite https://www.mdpi.com/1648-9144/60/11/1883
|
|---|
| Jezik: | Angleški jezik |
|---|
| Tipologija: | 1.01 - Izvirni znanstveni članek |
|---|
| Organizacija: | UKC LJ - Univerzitetni klinični center Ljubljana
|
|---|
| Povzetek: | Background and Objectives: This study aimed to investigate the risk and mechanisms of air entry into the extracorporeal membrane oxygenation (ECMO) circuit through the central venous catheter (CVC) in a veno-venous configuration. The primary goal was to assess the impact of different air volumes on ECMO circuit performance at varying pump speeds. Material and Methods: The study utilized a circuit model to simulate ECMO conditions and evaluate the potential entry points of air, specifically through the unprotected lumen of the CVC. Various interventions, such as the use of a closed three-way stopcock or clave, were implemented to assess their efficacy in preventing air entry. Results: The unprotected lumen of the central venous catheter posed a significant risk for air entry into the ECMO circuit. The introduction of a closed three-way stopcock or clave proved effective in preventing air ingress through the central venous catheter. Auditory cues, such as a distinct hissing sound, served as an early warning sign of air presence in the circuit. The study demonstrated that even small volumes of air, as minimal as 1 mL, could pass through the oxygenator at specific pump speeds, and larger volumes could lead to pump dysfunction. Conclusions: The study identified the unprotected lumen of the central venous catheter as a potential entry point for air into the ECMO circuit. The use of a closed three-way stopcock or one-way valve was found to be a reliable protective measure against air infiltration. Early detection through the observation of a hissing sound in the circuit provided a valuable warning sign. These findings contribute to enhancing the safety and performance of ECMO systems by minimizing the risk of air embolism. |
|---|
| Ključne besede: | air embolism, complications, extracorporeal membrane oxygenation, simulation |
|---|
| Status publikacije: | Objavljeno |
|---|
| Verzija publikacije: | Objavljena publikacija |
|---|
| Leto izida: | 2024 |
|---|
| Št. strani: | str. 1-8 |
|---|
| Številčenje: | Vol. 60, iss. 11, [article no.] 1883 |
|---|
| PID: | 20.500.12556/DiRROS-30059  |
|---|
| UDK: | 616.1 |
|---|
| ISSN pri članku: | 1648-9144 |
|---|
| DOI: | 10.3390/medicina60111883  |
|---|
| COBISS.SI-ID: | 241108227  |
|---|
| Opomba: |
Nasl. z nasl. zaslona;
Opis vira z dne 2. 7. 2025;
|
|---|
| Datum objave v DiRROS: | 12.06.2026 |
|---|
| Število ogledov: | 74 |
|---|
| Število prenosov: | 46 |
|---|
| Metapodatki: |  |
|---|
|
:
|
Kopiraj citat |
|---|
| | | | Objavi na: |  |
|---|
Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše
podrobnosti ali sproži prenos. |