Naslov:Virtual modelling of novel applicator prototypes for cervical cancer brachytherapy
Avtorji:ID Petrič, Primož (Avtor)
ID Hudej, Robert (Avtor)
ID Al-Hammadi, Noora (Avtor)
ID Šegedin, Barbara (Avtor)
Datoteke:.pdf PDF - Predstavitvena datoteka, prenos (671,88 KB)
MD5: FA0158B694908CD0ED4CFB40BB1B694A
 
Jezik:Angleški jezik
Tipologija:1.01 - Izvirni znanstveni članek
Organizacija:Logo OI - Onkološki inštitut Ljubljana
Povzetek:Standard applicators for cervical cancer Brachytherapy (BT) do not always achieve acceptable balance between target volume and normal tissue irradiation. We aimed to develop an innovative method of Targetvolume Density Mapping (TDM) for modelling of novel applicator prototypes with optimal coverage characteristics. Patients and methods. Development of Contour-Analysis Tool 2 (CAT-2) software for TDM generation was the core priority of our task group. Main requests regarding software functionalities were formulated and guided the coding process. Software validation and accuracy check was performed using phantom objects. Concepts and terms for standardized workflow of TDM post-processing and applicator development were introduced. Results. CAT-2 enables applicator-based co-registration of Digital Imaging and Communications in Medicine (DICOM) structures from a sample of cases, generating a TDM with pooled contours in applicator-eye-view. Each TDM voxel is assigned a value, corresponding to the number of target contours encompassing that voxel. Values are converted to grey levels and transformed to DICOM image, which is transported to the treatment planning system. Iso-density contours (IDC) are generated as lines, connecting voxels with same grey levels. Residual Volume at Risk (RVR) is created for each IDC as potential volume that could contain organs at risk. Finally, standard and prototype applicators are applied on the TDM and virtual dose planning is performed. Dose volume histogram (DVH) parameters are recorded for individual IDC and RVR delineations and characteristic curves generated. Optimal applicator configuration is determined in an iterative manner based on comparison of characteristic curves, virtual implant complexities and isodose distributions. Conclusions. Using the TDM approach, virtual applicator prototypes capable of conformal coverage of any target volume, can be modelled. Further systematic assessment, including studies on clinical feasibility, safety and effectiveness are needed before routine use of novel prototypes can be considered.
Ključne besede:cervical cancer, brachytherapy, applicators, virtual modelling
Status publikacije:Objavljeno
Verzija publikacije:Objavljena publikacija
Datum objave:01.12.2016
Založnik:Association of Radiology and Oncology
Leto izida:2016
Št. strani:str. 433-441, IX
Številčenje:Vol. 50, no. 4
Izvor:Ljubljana
PID:20.500.12556/DiRROS-18886 Novo okno
UDK:615.849
ISSN pri članku:1318-2099
DOI:10.1515/raon-2016-0048 Novo okno
COBISS.SI-ID:2567803 Novo okno
Avtorske pravice:by Authors
Datum objave v DiRROS:09.05.2024
Število ogledov:424
Število prenosov:177
Metapodatki:XML DC-XML DC-RDF
:
Kopiraj citat
  
Objavi na:Bookmark and Share


Postavite miškin kazalec na naslov za izpis povzetka. Klik na naslov izpiše podrobnosti ali sproži prenos.

Gradivo je del revije

Naslov:Radiology and oncology
Skrajšan naslov:Radiol. oncol.
Založnik:Slovenian Medical Society - Section of Radiology, Croatian Medical Association - Croatian Society of Radiology
ISSN:1318-2099
COBISS.SI-ID:32649472 Novo okno

Sekundarni jezik

Jezik:Slovenski jezik
Ključne besede:maternični vrat, rak (medicina), brahiterapija, aplikatorji


Nazaj