1.
Mechanisms of premature fracture in modular neck stems made of CoCrMo/Ti6Al4V and Ti6Al4V/Ti6Al4V alloy
Drago Dolinar, Miro Gorenšek, Klemen Avsec, Barbara Šetina, Matej Hočevar, Matjaž Godec, Borut Žužek, Mojca Debeljak, Monika Jenko, John T. Grant, Boštjan Kocjančič, 2023, izvirni znanstveni članek

Povzetek: In this paper, we present the mechanisms of premature fracture of modular neck stems in two case studies: (I) when the neck and stem are both made of the same Ti6Al4V alloy, and (II) when the neck and stem are made from two different alloys, CoCrMo and Ti6Al4V alloy. Our study integrates two orthopedic patients who have undergone primary uncemented THA for usual indications in two orthopedic centers (Community Health Centre and University Medical Centre). Both centers are part of the national public health care system. Both surgeries were performed by two skilled orthopedic surgeons with more than 10 years of experience in THA. The survivorship of the modular neck of cast CoCrMo alloy was 24 months. The survivorship of the modular neck from Ti6Al4V alloy was 84 months. Multivariate analyses were performed to assess the differences in the fretting, corrosion, and fatigue of the two prematurely failed modular neck stems: stereo light microscopy (SLM), scanning electron microscopy (SEM), X-ray energy-dispersive spectroscopy (EDS), and electron backscatter diffraction (EBSD). Patient demographic information, including sex, age, body mass index, survivorship of implants, and reason for the revision, was collected from medical records. We found that fretting and fatigue occurred on both neck-stem retrievals due to additional galvanic corrosion, but the CoCrMo/Ti6Al4V alloy system suffered more corrosion due to additional galvanic corrosion and fractured earlier than the Ti6Al4V/Ti6Al4V metal alloy system. Both metallic alloy systems used in this application are known to be highly corrosion-resistant, but the bio-tribo-corrosion processes need to be understood in detail and characterized so that appropriate improvements in design and materials can be made.
Ključne besede: total hip arthroplasty, modular neck, Ti6Al4V alloy, CoCrMo alloy, corrosion
Objavljeno v DiRROS: 07.02.2024; Ogledov: 229; Prenosov: 125
.pdf Celotno besedilo (5,12 MB)
Gradivo ima več datotek! Več...

2.
3.
Tribocorrosive study of new and in vivo exposed nickel titanium and stainless steel orthodontic archwires
Tadeja Kosec, Petra Močnik, Uroš Mezeg, Andraž Legat, Maja Ovsenik, Monika Jenko, John T. Grant, Jasmina Primožič, 2020, izvirni znanstveni članek

Povzetek: The surface, corrosion and wear properties of new and in vivo exposed nickel titanium (NiTi) and stainless steel (SS) archwires used in orthodontic treatment were investigated. Electrochemical and tribo-electrochemical tests in artificial saliva were performed in order to define corrosion properties and to estimate wear rate of new and in vivo exposed NiTi and SS archwires. The surface chemical analysis of the passive film on the NiTi and SS archwires before and after tribocorrosion tests was performed by Auger Electron Spectroscopy (AES). In vivo exposed NiTi and SS archwires had better electrochemical properties than new archwires due to the protective nature of oral deposits. Total wear and coefficients of friction were higher among in vivo exposed archwires and higher in NiTi archwires in comparison to SS archwires. The estimated thickness of the TiO2 passive film on as-received NiTi is 8 nm, while the passive Cr2O3 film on as-received SS is just 1–2 nm. On in vivo exposed NiTi archwire, a 60–80 nm thick organic film/dental plaque was observed, and on SS, it was thinner, at about 60 nm. This research shows the importance of combining AES with electrochemical testing, to characterize tribocorrosive properties of NiTi and SS orthodontic archwires.
Ključne besede: archwires, NiTi, stainless steel, wear
Objavljeno v DiRROS: 24.08.2023; Ogledov: 265; Prenosov: 114