Zagotavljanje kakovosti v ultrazvočni diagnostiki
Authors:
ID
Rainer, Sašo
(Author)
Files:
PDF - Presentation file,
download
(225,33 KB)
MD5: 26B30AA3216B3FF27F21E63D75DB4ED4
Language:
Slovenian
Typology:
1.02 - Review Article
Organization:
OI - Institute of Oncology
Abstract:
Izvajanje programa za zagotavljenje kakovosti ima velik pomen tudi v ultrazvočni diagnostiki, saj lahko poslabšanje kakovosti slike in nekaterih funkcij pomembno vpliva na zmožnost odkrivanja in pravilne opredelitve patoloških procesov. Elementi programa za zagotavljanje kvalitete so: pravilnoravnanje z aparaturo, njeno dnevno vzdrževanje, detekcija slabšanja kakovosti pri dnevnem delu in redno testiranje vsakih šest mesecev. Testiranjezajema vizualno inspekcijo aparature (sond, mehaničnih delov, filtra, kablov in kontaktov), oceno kakovosti dokumentacijskih posnetkov, testiranje s fantomom in dokumentacijo rezultatov. S fantomom preverjamo specifične parametre, ki neposredno zadevajo kakovost slike. V prispevku so okvirno obrazloženi ti parametri ter njihov pomen, opisane so tudi vrste ultrazvočnih fantomov. Redno izvajanje programa za zagotavljanje kakovosti bi moralo postati obvezen sestavni del aktivnosti vsake ultrazvočne diagnostične enote.
Publication status:
Published
Publication version:
Version of Record
Publication date:
01.01.1999
Publisher:
Association of Radiology and Oncology
Year of publishing:
1999
Number of pages:
str. S42-6
Numbering:
Letn. 33, suppl. 2
Source:
Ljubljana
PID:
20.500.12556/DiRROS-17488
UDC:
616-006
ISSN on article:
1318-2099
COBISS.SI-ID:
10738137
Copyright:
by Authors
Note:
Zagotavljanje kakovosti v slikovni diagnostiki dojke; BSDOCID48374;
Publication date in DiRROS:
08.12.2023
Views:
536
Downloads:
78
Metadata:
Cite this work
Plain text
BibTeX
EndNote XML
EndNote/Refer
RIS
ABNT
ACM Ref
AMA
APA
Chicago 17th Author-Date
Harvard
IEEE
ISO 690
MLA
Vancouver
:
RAINER, Sašo, 1999, Zagotavljanje kakovosti v ultrazvočni diagnostiki.
Radiology and oncology
[online]. 1999. Vol. 33, no. suppl. 2, p. s42-6. [Accessed 20 April 2025]. Retrieved from: https://dirros.openscience.si/IzpisGradiva.php?lang=eng&id=17488
Copy citation
Share:
Searching for similar works...
Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.
Record is a part of a journal
Title:
Radiology and oncology
Shortened title:
Radiol. oncol.
Publisher:
Slovenian Medical Society - Section of Radiology, Croatian Medical Association - Croatian Society of Radiology
ISSN:
1318-2099
COBISS.SI-ID:
32649472
Secondary language
Language:
English
Title:
Quality assurance in ultrasound
Abstract:
Quality assurance is essential also in ultrasonography; image degradetion and malfunction of the equipment can adversely influence the ability of the systemto detect and correctly carracterize pathological processes. Quality assurance program consist of optimal operatioon and daily care of imaging equipment and docentation devices, detection of image quality degradation during routine equipment operation, and periodic quality assurance tests. Testing comprises of visual inspection of the equipment (transducers, connection sockets and cables, mechanical components, filters), hard copy evaluation, phantom testing and documentation of the results. With phantom testing, specific parameters related to image quality are evaluated. These parameters are basicaly outlined, as well as the types of ultrasound quality assurance phantoms. Periodic quality assurance testing should become the integral part of routine activity in every diagnostic ultrasound unit.